A J Malyadri
Articles written in Pramana – Journal of Physics
Volume 57 Issue 1 July 2001 pp 215-218 Contributed Papers, Others
The automatic liquid nitrogen filling system for GDA detectors
Rakesh Kumar A J Malyadri S Muralithar Ruby Shanti S K Saini Kusum Rani B P Ajith Kumar Rajesh Kumar R K Bhowmik
An indigenously developed automatic liquid nitrogen (LN2) filling system has been installed in gamma detector array (GDA) facility at Nuclear Science Centre. Electro-pneumatic valves are used for filling the liquid nitrogen into the high purity germanium detector cryostat. The temperature of the out-flowing gas/liquid from the cryostat is monitored using platinum resistor thermometer. The program allows for automatic filling at regular intervals with temperature monitoring from a remote terminal.