    • Keywords

       

      Microchannel plate; X-ray framing; high voltage pulser

    • Abstract

       

      A single-frame X-ray framing camera has been set up for fast imaging of X-ray emissions from pulsed plasma sources. It consists of two parts, viz. an X-ray pin-hole camera using an open-ended microchannel plate (MCP) detector coupled to a CCD camera, and a high voltage short duration gate pulse for the MCP. The camera uses a 10-Μm pin-hole aperture for imaging on the MCP detector with a magnification of 6 X. The high voltage pulser circuit generates a pulse of variable duration from 5 to 30 ns (at 70% of peak amplitude) with variable amplitude from 800 V to 1.25 kV, and is triggered through a laser pulse synchronized with the event to be recorded. The performance of the system has been checked by recording X-ray emission from a laser-produced copper plasma. A reduction factor of ∼ 6.5 is seen in the dark current contribution as the MCP gate pulse is decreased from 250Μs to 5 ns duration.

    • Author Affiliations

       

      J Upadhyay1 J A Chakera1 C P Navathe1 P A Naik1 A S Joshi1 P D Gupta1

      1. Laser Plasma Division, Centre for Advanced Technology, P O CAT, Indore - 452013