      Terminal pumping system; AMS; Pelletron; electrostatic accelerator; accelerator mass spectrometry

    • Abstract

       

      A recirculating terminal pumping system has been installed in the 3 MV Pelletron tandem accelerator in Lund. An extremly limited space in the high voltage terminal and the absence of electrical power in the terminal, forced us to provide a unique design for the installation and powering of the new pumps. Details of the technical design, as well as experience of the use of the new system for accelerator mass spectrometry, will be given.

    • Author Affiliations

       

      R Hellborg1 K Hākansson1 M Faarinen1 M Kiisk1 P Persson1 G Skog1 K Stenström1