• Fulltext

       

        Click here to view fulltext PDF


      Permanent link:
      https://www.ias.ac.in/article/fulltext/pram/055/05-06/0733-0740

    • Keywords

       

      Thomson scattering; plasma diagnostics

    • Abstract

       

      Thomson scattering technique based on high power laser has already proved its superoirity in measuring the electron temperature (Te and density (ne) in fusion plasma devices like tokamaks. The method is a direct and unambiguous one, widely used for the localised and simultaneous measurements of the above parameters. In Thomson scattering experiment, the light scattered by the plasma electrons is used for the measurements. The plasma electron temperature is measured from the Doppler shifted scattered spectrum and density from the total scattered intensity. A single point Thomson scattering system involving a Q-switched ruby laser and PMTs as the detector is deployed in ADITYA tokamak to give the plasma electron parameters. The system is capable of providing the parameters Te from 30 eV to 1 keV and ne from 5 × 1012cm−3−5 × 1013cm−3. The system is also able to give the parameter profile from the plasma center (Z=0 cm) to a vertical position of Z=+22 cm to Z=−14 cm, with a spatial resolution of 1 cm on shot to shot basis. This paper discusses the initial measurements of the plasma temperature from ADITYA.

    • Author Affiliations

       

      R Rajesh1 B Ramesh Kumar1 S K Varshney1 Manoj Kumar1 Chhaya Chavda1 Aruna Thakkar1 N C Patel1 Ajai Kumar1 Aditya Team1

      1. Institute for Plasma Research, Bhat, Gandhinagar - 382 428, India
    • Dates

       
  • Pramana – Journal of Physics | News

    • Editorial Note on Continuous Article Publication

      Posted on July 25, 2019

      Click here for Editorial Note on CAP Mode

© 2022-2023 Indian Academy of Sciences, Bengaluru.