An automated thermal relaxation calorimeter for operation at low temperature (0.5 K<T<10 K)
S Banerjee M W J Prins K P Rajeev A K Raychaudhuri
Click here to view fulltext PDF
Permanent link:
https://www.ias.ac.in/article/fulltext/pram/039/04/0391-0404
We describe an automated calorimeter for measurement of specific heat in the temperature range 10 K>T>0.5 K. It uses sample of moderate size (100–1000 mg), has a moderate precision and accuracy (2%–5%), is easy to operate and the measurements can be done quickly with He4 economy. The accuracy of this calorimeter was checked by measurement of specific heat of copper and that of aluminium near its superconducting transition temperature.
S Banerjee1 M W J Prins1 K P Rajeev1 A K Raychaudhuri1
Volume 97, 2023
All articles
Continuous Article Publishing mode
Click here for Editorial Note on CAP Mode
© 2022-2023 Indian Academy of Sciences, Bengaluru.