• Plastics as nuclear track detectors for thermal neutron dosimetry

    • Fulltext

       

        Click here to view fulltext PDF


      Permanent link:
      https://www.ias.ac.in/article/fulltext/pram/020/06/0559-0567

    • Keywords

       

      Solid state nuclear track detectors; neutron dosimeter; fluence sensitivity; dose sensitivity; spark counter; electrochemical etching

    • Abstract

       

      An attempt is made to determine the response of CR-39 and cellulose nitrate plastic track detectors subjected to thermal neutrons. The α-particles are produced from (n, α) reactions in lithium tetraborate convertor placed in contact with different plastics and are recorded in the detectors. The corrected track density gives a fluence sensitivity and dose sensitivity of the order of 10−4 tracks per neutron and 102 tracks/cm2 mrem respectively. A linear relationship is observed between track density and neutron fluence.