• Fulltext

       

        Click here to view fulltext PDF


      Permanent link:
      https://www.ias.ac.in/article/fulltext/jcsc/133/0046

    • Keywords

       

      Polyaniline; 4-chloroaniline; graphitized mesoporous carbon; redox polymer; electropolymerization; electrochemical pH sensor.

    • Abstract

       

      Development of a new pH sensor system, which is simple to prepare, sensitive, selective and workable with low volume, is demanding research in biomedical and environmental studies. In the literature,organic molecules like methylene blue, toluidine blue, hydroquinone, catechol, anthraquinone and polyaniline-based redox probes have been widely used for this purpose. In general, these redox probes have easilyinterfered with common biochemicals such as dopamine, ascorbic acid, NADH, H2O2, cysteine, hydrazine,and some transition metal ions, etc., and in turn to marked potential and current drifts (pH-false positiveresponse). In this work, a highly redox-active, stable and interference-free redox polymer based on poly(4-chloroaniline) (PANI(4-Cl)) modified graphitized mesoporous carbon (GMC), designated as GMC@PANI(4-Cl), has been prepared using 4-chloroaniline as a monomer in pH 7 phosphate buffer solution. The new redox polymer system showed a distinct redox peak at Eo’= 0.15 V vs Ag/AgCl with a stable voltammetric response. Transmission electron microscope analysis of the redox polymer composite had shown adhesion of black polymeric solid due to polyaniline like the molecular system as a surface layer on the GMC material.The constructed calibration plot was linear in the pH window 2-11 with a slope and regression values - 58 mV pH-1 and 0.9997, respectively. The GMC@PANI(4-Cl) modified electrode showed a sensitive andselective pH monitoring without any interference from the common biochemicals as listed above. As apractical application, pH sensing of commercial pH solutions, undiluted urine and saliva samples weredemonstrated. Also, a three-in-one screen-printed carbon modified GMC@PANI(4-Cl) was explored for pHmonitoring of a bacterial (E.coli) growth, which showed a comparable response with the conventional pHelectrode.

    • Graphical Abstract

       

    • Author Affiliations

       

      SAIRAMAN SAIKRITHIKA1 2 ANNAMALAI SENTHIL KUMAR1 2

      1. Nano and Bioelectrochemistry Research Laboratory, Carbon dioxide and Green Technology Research Centre, Vellore Institute of Technology University, Tamil Nadu, Vellore 632 014, India
      2. Department of Chemistry, School of Advanced Sciences, Vellore Institute of Technology, Tamil Nadu, Vellore 632 014, India
    • Dates

       
  • Journal of Chemical Sciences | News

    • Editorial Note on Continuous Article Publication

      Posted on July 25, 2019

      Click here for Editorial Note on CAP Mode

© 2021-2022 Indian Academy of Sciences, Bengaluru.