• Benzoquinone-mediated enzyme biosensor for amperometric determination of glucose

    • Fulltext

       

        Click here to view fulltext PDF


      Permanent link:
      https://www.ias.ac.in/article/fulltext/jcsc/109/04/0295-0301

    • Keywords

       

      Benzoquinone; amperometric biosensors; glucose oxidase; enzyme sensor; cyclic voltammetry

    • Abstract

       

      An amperometric enzyme sensor for the analysis of glucose is described. The sensor uses benzoquinone as a mediator for electron transfer between immobilized glucose oxidase and a pyrolytic graphite electrode. A linear current response, proportional to glucose concentration in the range 1–50mM, is observed. The influence of oxygen,pH and operating potential on the sensor performance is examined.

    • Author Affiliations

       

      A S N Murthy1 Jyotsna Sharma1

      1. Department of Chemistry, Indian Institute of Technology, New Delhi - 110 016, India
    • Dates

       
  • Journal of Chemical Sciences | News