• Fulltext

       

        Click here to view fulltext PDF


      Permanent link:
      https://www.ias.ac.in/article/fulltext/boms/008/05/0591-0595

    • Keywords

       

      Low temperature diffraction; diffractometer cryo-assembly; helium gas cryo-tip; closed-cycle cryogenerator

    • Abstract

       

      A single-stage closed-cycle helium gas cryogenerator, capable of cooling between 77 K and 300 K, has been installed on the 4-circle neutron diffractometer at CIRUS. The cryo-tip assembly has been incorporated directly on the ϕ-circle of the full-circle crystal orienter in place of the goniometer head used in room temperature experiments. The salient features of this low-temperature attachment and its performance are discussed.

    • Author Affiliations

       

      A Sequeira1 Sandhya Bhakay-Tamhane1 S R Sawant1 2

      1. Neutron Physics Division, Bhabha Atomic Research Centre, Trombay, Bombay - 400 085, India
      2. Physics Group Workshop, Bhabha Atomic Research Centre, Trombay, Bombay - 400 085, India