• Sophisticated equipment developed for growth and evaluation of perfection of nearly perfect crystals

    • Fulltext

       

        Click here to view fulltext PDF


      Permanent link:
      https://www.ias.ac.in/article/fulltext/boms/006/02/0129-0149

    • Keywords

       

      Crystal puller; microfocus x-ray generator; x-ray diffraction topography camera; x-ray diffractometer

    • Abstract

       

      This paper reviews four major equipment developed at the National Physical Laboratory for growth and perfection evaluation of single crystals, namely (i) a crystal puller for growth of nearly perfect crystals by the Czochralski method; (ii) a microfocus x-ray generator; (iii) an x-ray diffraction topography camera; and (iv) a triple crystal x-ray diffractometer. The crystal puller can provide smooth, uniform and variable pulling rates. The maximum length of pull is nearly 60cm. Efforts have been made to isolate vibrations. Nearly perfect single crystals of KCl, KBr and NaCl with maximum diameter of ∼ 60 mm have been grown. The crystals give diffraction curves with half width in the range of 10–30 sec of arc. In the projection topographs, dislocations can be resolved and characterized. The microfocus x-ray generator is a demountable continuously evacuated system with specially designed electron gun and anode assembly. The vacuum is continuously monitored for ease of maintenance. In the point focus mode the spot size is 40 µm on the anode. X-ray topography system is a versatile equipment used for projection and section topography. It can provide 360° rotations to the specimen disc around an axis perpendicular to it. Rotations of a few sec of arc can be given to the specimen around a vertical axis. Typical diffraction curves of a dislocation-free crystal and a crystal with boundaries are shown. Well-resolved images of dislocations are shown in a topograph as an illustration. In the triple crystal x-ray diffractometer a highly collimated and monochromated Kα1 exploring x-ray beam is obtained by combining microfocus source, a special collimator and crystal monochromators of Bonse-Hart type. With this beam very narrow diffraction curves with half width of about 5 sec of arc can be recorded. Typical results of measurement of diffuse x-ray scattering (dxs) on nearly perfect silicon single crystals are discussed. It has been observed that the contribution of phonons to thedxs is negligible. Thedxs is mainly due to point defect aggregates.

    • Author Affiliations

       

      Krishan Lal1 Ajit Ram Verma1

      1. National Physical Laboratory, Hillside Road, New Delhi - 110 012, India
    • Dates

       
  • Bulletin of Materials Science | News

    • Dr Shanti Swarup Bhatnagar for Science and Technology

      Posted on October 12, 2020

      Prof. Subi Jacob George — Jawaharlal Nehru Centre for Advanced Scientific Research, Jakkur, Bengaluru
      Chemical Sciences 2020

      Prof. Surajit Dhara — School of Physics, University of Hyderabad, Hyderabad
      Physical Sciences 2020

    • Editorial Note on Continuous Article Publication

      Posted on July 25, 2019

      Click here for Editorial Note on CAP Mode

© 2023-2024 Indian Academy of Sciences, Bengaluru.